    摘要:

    目的对某品牌紫外线A/B型坐便器盖板紫外线灯的杀菌和消毒效果进行检测。方法采用载体定量杀菌试验,按照产品说明书要求,分别开启紫外线灯15、20、25、30、35 min作用于坐板载体上的金黄色葡萄球菌、大肠埃希菌、白假丝酵母菌、枯草杆菌黑色变种芽孢;现场消毒试验,在开启紫外线灯前和照射预定时间后用棉拭子涂抹坐板采样,进行活菌培养计数。结果载体定量杀菌试验,金黄色葡萄球菌、大肠埃希菌、白假丝酵母菌、枯草杆菌黑色变种芽孢的平均杀灭对数值均≥3.00。现场消毒试验的30份受试样品平均杀灭对数值为3.57,各次杀灭自然菌的对数值均>1.00。结论该品牌紫外线A/B型坐便器盖板紫外线灯的消毒杀菌效果合格。

    Abstract:

    ObjectiveTo detect the disinfection and sterilization effect of  A/B type ultraviolet (UV) lamp on closestool. MethodsCarrier quantitative germicidal test  was performed, according to explaination of the  product, UV lamp irradiated for 15, 20, 25, 30, 35 minutes respectively on Staphylococcus aureus, Escherichia coli, Candida albicans and Bacillus subtilis on the carriers; onthespot disinfection test was also performed, before and after turning on the VU lamp for scheduled  time, samples were taken with cotton swab smear method, live bacterial counts were calculated.ResultsCarrier quantitative germicidal test showed that the average killing logarithm value of Staphylococcus aureus, Escherichia coli, Candida albicans and Bacillus subtilis were all≥3.00.  Onthespot disinfection test of 30 samples showed that the average killing logarithm value was 3.57, killing logarithm value of natural bacteria were all >1.00. ConclusionThe disinfection and sterilization effect of the  UV lamp on the A/B  type  closestool  is qualified.

    参考文献
    相似文献
引用本文

龚玉姣,肖扬,陈建东,等.坐便器盖板紫外线灯消毒杀菌效果研究[J]. 中国感染控制杂志,2009,8(3):181-183.
GONG Yujiao, XIAO Yang, CHEN Jiandong, et al. Disinfection and sterilization effect of  ultraviolet lamp on  closestool[J]. Chin J Infect Control, 2009,8(3):181-183.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2008-09-16
  • 最后修改日期:2008-11-29
  • 录用日期:
  • 在线发布日期: 2009-05-30
  • 出版日期: