292所医疗机构内镜自动清洗消毒机使用及管理现状调查
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

R187

基金项目:

江西省科技厅自然科学基金资助项目(20212BAB206023)


Application and management status of automated endoscope reprocessors in 292 medical institutions
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 文章评论
    摘要:

    目的 了解中国内镜自动清洗消毒机(AER)使用、管理及维护现状。方法 采用便利抽样法,对292所医疗机构消化内镜中心AER使用、管理、维护、监测,以及人员培训等方面进行问卷调查。结果 共发放问卷316份,有效回收292份,有效回收率92.41%。292所医疗机构中198所(67.81%)配备AER,其中三级医疗机构AER配备率为75.11%,二级医疗机构AER配备率为41.27%。88.38%的医疗机构在内镜放入AER之前进行手工清洗,但仍有2.02%的医疗机构未执行该步骤,主要集中在三级医疗机构。AER使用中最常见的问题是连接管脱落或破损(83.33%),其次是AER漏消毒剂(43.94%)、控制面板失灵(43.94%) 和AER管路不出水或出水不畅(42.93%)。92.42%的医疗机构定期对AER进行自身消毒,72.73%的医疗机构定期更换AER空气过滤器, 80.30%的医疗机构定期更换AER水过滤器。96.46%的医疗机构每日使用AER之前进行消毒剂浓度监测,87.88%的医疗机构定期监测AER终末漂洗水,96.97%的医疗机构对新入职人员首次使用AER进行了专业培训。结论 中国各级医疗机构AER配备率有待提升,大部分医疗机构能够较好地监测和维护AER,但在使用过程中仍存在诸多问题。三级医疗机构因内镜诊疗量大,手工清洗执行率不足,需优化工作流程并加强培训。二级医疗机构则面临设备短缺、监测执行率低等挑战,需加大资金投入和监管力度。建议生产制造商不断优化AER设计,医疗机构建立AER标准化使用流程,相关部门完善AER使用及管理规范,以提升内镜清洗消毒质量,保障患者诊疗安全。

    Abstract:

    Objective To understand the current status of application, management, and maintenance of automated endoscope reprocessors (AERs) in China. Methods A questionnaire survey was conducted using convenience sampling method on the application, management, maintenance, and monitoring of AERs, as well as personnel training in digestive endoscopy centers of 292 medical institutions (MIs). Results A total of 316 questionnaires were distributed, and 292 were available, with an effective response rate of 92.41%. Among the 292 MIs, 198 (67.81%) were equipped with AERs, with equipment rates of 75.11% in tertiary MIs and 41.27% in secondary MIs. 88.38% of MIs performed manual cleaning before placing endoscopes into AERs, while 2.02% of MIs still didn’t perform this step, mainly in tertiary MIs. The most common problems in the application of AERs were the detachment or damage of connecting pipes (83.33%), followed by disinfectant leakage of AERs (43.94%), malfunction of control panel (43.94%), and failure or poor water flow of AER pipeline (42.93%). 92.42% of MIs regularly disinfected AERs, 72.73% of MIs regularly replaced AER air filters, and 80.30% of MIs regularly replaced AER water filters. 96.46% of MIs monitored the concentration of disinfectants before daily application of AERs, 87.88% of MIs regularly monitored the final rinse water of AERs, and 96.97% of MIs provided professio-nal training for new employees to use AERs for the first time. Conclusion The equipping rate of AERs at all levels of MIs in China still needs to be improved. Most MIs are able to monitor and maintain AERs well, but there are still many problems in the application process. Due to the large volume of endoscopic diagnosis and treatment as well as insufficient execution rate of manual cleaning in tertiary MIs, it is necessary to optimize workflow and strengthen training. Secondary MIs face challenges such as equipment shortages and low monitoring execution rates, requiring increased funding and regulatory efforts. It is recommended that manufacturers continuously optimize AER design, MIs establish standardized AER application protocol, and relevant departments improve AER application and ma-nagement standards, in order to improve the quality of endoscopic cleaning and disinfection and ensure patient safety during diagnosis and treatment.

    参考文献
    相似文献
引用本文

甄静,陈刘吉,刘海霞,等.292所医疗机构内镜自动清洗消毒机使用及管理现状调查[J]. 中国感染控制杂志,2025,24(9):1278-1285. DOI:10.12138/j. issn.1671-9638.20252060.
ZHEN Jin, CHEN Liuji, LIU Haixia, et al. Application and management status of automated endoscope reprocessors in 292 medical institutions[J]. Chin J Infect Control, 2025,24(9):1278-1285. DOI:10.12138/j. issn.1671-9638.20252060.

复制
分享
文章指标
  • 摘要阅读次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2025-01-21
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2025-09-23
  • 出版日期: