Investigation and disposal of bacterial contamination in pure water system during acceptance of newly built endoscopy center
Author:
Affiliation:

1.Department of Healthcare-associated Infection Management, Zhengzhou People's Hospital, Zhengzhou 450002, China;2.Department of Logistics, Zhengzhou People's Hospital, Zhengzhou 450002, China;3.Department of Laboratory Medicine, Zhengzhou People's Hospital, Zhengzhou 450002, China;4.Department of Medical Equipment, Zhengzhou People's Hospital, Zhengzhou 450002, China

Clc Number:

+4]]>

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Comments
    Abstract:

    Objective To investigate the causes for number of bacterial colonies exceeding the standard in the pure water system found during the acceptance of newly built endoscopy center, and put forward solutions. Methods Water specimens were collected from all steps of pure water preparation and use, the causes for number of bacterial colonies exceeding the standard were identified, disposal measures according to the causes were formulated, and the maintenance time of intensive disinfection effect was monitored. Results The investigation result inferred that the newly laid water pipeline between the outlet of pure water tank and inlet of the endoscope before entering the endoscopic decontamination equipment is the main cause for the bacteria exceeding the standard. According to the causes, combined disposal measures such as pipeline transformation, adding disinfection facilities, cleaning and disinfection of water pipeline and decontamination equipment are formulated. After comprehensive intervention, no bacteria were detected at each sampling point, the endoscopy center was successfully put into use, and the interval of waterline intensive disinfection was determined to be 12 weeks. Conclusion The acceptance before operation of the newly built endoscope center is very important, the new pure water pipeline needs to be flushed before laying to ensure the quality of final rinsing water.

    Reference
    Related
Get Citation

张枭然,阎颖,千新玲,等.新建内镜中心验收时纯水系统染菌情况调查与处置[J].中国感染控制杂志英文版,2021,(12):1114-1118. DOI:10.12138/j. issn.1671-9638.20218313.
Xiao-ran ZHANG, Ying YAN, Xin-ling QIAN, et al. Investigation and disposal of bacterial contamination in pure water system during acceptance of newly built endoscopy center[J]. Chin J Infect Control, 2021,(12):1114-1118. DOI:10.12138/j. issn.1671-9638.20218313.

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:April 07,2021
  • Revised:
  • Adopted:
  • Online: April 26,2024
  • Published: