Enteroscope secondary contamination caused by compressed air for drying: one case report
Author:
Affiliation:

Department of Healthcare-associated Infection Management, Fujian Medical University Union Hospital, Fuzhou 350001, China

Clc Number:

R472.1

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Comments
    Abstract:

    Objective To analyze the causes of exceeding standard of colony count of enteroscope after disinfection, and further improve the reprocessing level of flexible endoscope. Methods Data of the third quarter routine environmental hygiene monitoring of digestive endoscopy center of a hospital conducted in August 2021 were analyzed, causes of secondary contamination of enteroscope after disinfection were found out and rectified. Results The monitored four gastroscopes, two specimens of purified water, two specimens of in-use disinfectants, three specimens of environmental object surfaces and two specimens of hygienic hand disinfection of health care workers were all qualified; however, a large number of bacteria were detected from four enteroscopies, bacterial colony number were 252, 315, 273, and 268 CFU/piece respectively, all specimens were detected Rhizobium radiobacter, Micrococcus luteus, and Sphingomonas paucimobilis. Possible risk factors leading to the failure of enteroscope monitoring were identified, specimens of air from air spray gun, inside of air filter and air line of filter outlet were detected further, a large number of the above microorganisms were found out.After the replacement of air compressor, air filter and air line, the number of bacteria in air from air spray gun decreased obviously.The 4 unqualified enteroscopes were cleaned, disinfected and dried again, bacterial contamination was detected, all were all qualified. Conclusion Contamination of air line of the digestive endoscope cleaning workstation is the main cause for the excessive number of bacteria in the lumen of enteroscope after disinfection.It is suggested that air line maintenance and biological monitoring should be included in the corresponding guidelines, so as to reduce healthcare-associated infection.

    Reference
    Related
Get Citation

聂渝莉,张巧禹,徐秀莉,等.一起干燥用压缩空气引起的肠镜二次污染案例报告[J].中国感染控制杂志英文版,2022,(9):911-915. DOI:10.12138/j. issn.1671-9638.20222932.
Yu-li NIE, Qiao-yu ZHANG, Xiu-li XU, et al. Enteroscope secondary contamination caused by compressed air for drying: one case report[J]. Chin J Infect Control, 2022,(9):911-915. DOI:10.12138/j. issn.1671-9638.20222932.

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:May 24,2022
  • Revised:
  • Adopted:
  • Online: April 28,2024
  • Published: